dc.description.abstract | James One (MindProber Labs) is a wireless psychophysiological device comprising two sensors: one measuring electrodermal activity (EDA), the other photoplethysmography (PPG). This paper reports the validation of James One's EDA sensor by comparing its signal against a research grade polygraph. Twenty participants were instructed to perform breathing exercises to elicit the modulation of EDA and heart rate, while the physiological signal was captured simultaneously on James One and a Biopac MP36. The resulting EDA and PPG records collected from both systems were comprehensively compared. Results suggest that James One captures EDA signal with a quality comparable to a research grade equipment, this constituting a reliable means of capturing data while minimizing setup time and intrusiveness. | por |
dc.description.sponsorship | P.S.M. was supported by an FCT fellowship grant (PhD-iHES program) with the reference PDE/BDE/113601/2015. | por |
dc.language.iso | eng | por |
dc.publisher | MDPI | por |
dc.relation | PDE/BDE/113601/2015 | por |
dc.rights | openAccess | por |
dc.subject | Adult | por |
dc.subject | Algorithms | por |
dc.subject | Female | por |
dc.subject | Heart Rate | por |
dc.subject | Humans | por |
dc.subject | Male | por |
dc.subject | Photoplethysmography | por |
dc.subject | Psychophysiology | por |
dc.subject | Reproducibility of Results | por |
dc.subject | Signal Processing, Computer-Assisted | por |
dc.subject | Wireless Technology | por |
dc.subject | Electrodermal activity | por |
dc.title | Validation of wireless sensors for psychophysiological studies | por |
dc.type | article | por |
dc.peerreviewed | yes | por |
oaire.citationIssue | 22 | por |
oaire.citationVolume | 19 | por |
dc.identifier.doi | 10.3390/s19224824 | por |
dc.subject.wos | Science & Technology | por |
sdum.journal | Sensors | por |
Aparece nas coleções: | ICVS - Artigos em revistas internacionais / Papers in international journals |
Ficheiros deste registo:
Ficheiro | Descrição | Tamanho | Formato | |
---|---|---|---|---|
Silva-moreira-2019-Validation-of-wireless-sensors-for-.pdf | 17,88 MB | Adobe PDF | Ver/Abrir |