Título: | Mass calibration and Relative Humidity compensation requirements for optical portable particulate matter monitors: The IMPASHS (Impact of smoke-free policies in EU Member States) WP2 preliminary results |
Autor(es): | Ruprecht, Ario A. De Marco, Cinzia Boffi, Roberto Mazza, Roberto Lopez, Maria José Moshammer, Hanns Dautzenberg, Bertrand Clancy, Luke Precioso, José Invernizzi, Giovanny |
Palavras-chave: | Smoking Second-hand smoke Monitoring |
Data: | 2011 |
Editora: | Lippincott, Williams & Wilkins |
Revista: | Epidemiology |
Resumo(s): | Better knowledge of particulate matter (PM) concentrations needs portable, reliable, user friendly, low cost, real time mass analyzers of PM2.5 and PM10. Optical Particle Counters (OPC) measuring mass have manufacturer calibration specific gravity “K” factor referred to polystyrene latex particles which are completely different than those of the real world, therefore they require specific calibrations. Measurements are also subject to Relative Humidity (RH) heavy interference. |